基于双波长强度调制的法布里-珀罗滤光器中心波长定标 | |
其他题名 | Center Wavelength Calibration of Fabry‒Perot Filter Based on Dual-Wavelength Intensity Modulation |
朱瑾1,2![]() ![]() ![]() | |
发表期刊 | 光学学报/Acta Optica Sinica
![]() |
2024-04 | |
卷号 | 44期号:24页码:185-192 |
分类号 | P111.41 |
产权排序 | 第1完成单位 |
收录类别 | EI ; 核心 ; CSCD |
关键词 | Fabry-Perot滤光器 定标 腔长 中心波长 双波长强度调制 |
摘要 | 法布里-珀罗(F-P)滤光器是太阳二维成像光谱观测系统的重要组成部分,为了实现高精度太阳二维成像光谱观测,要求F-P滤光器必须具有准确的中心波长。由于滤光器只能提供控制器数值,无法直接获取波长信息,提出一种基于双波长强度调制的定标方法,利用F-P滤光器强度调制的周期性和双波长强度调制的差异性,建立控制器数值和滤光器腔长及波长的对应关系,实现波长精确定标。在实验室开展F-P滤光器定标测试,测试结果表明,对于Hα谱线,滤光器的波长定标精度达到0.0093?,满足1 m新真空太阳望远镜(NVST)下一代二维成像光谱观测系统的指标要求。 |
其他摘要 | Objective The Fabry‒Perot (FP) filter, with the advantages of narrow bandwidth, high transmittance, and large aperture, is an essential component in solar twodimensional imaging spectrometers. It is also a crucial optical element in the nextgeneration twodimensional imaging spectrometer for the New Vacuum Solar Telescope (NVST). To achieve highprecision solar spectrum observation, the FP filter must have an accurate center wavelength. However, in practical applications, the filter can only provide controller values and cannot obtain wavelength information directly. Additionally, the accuracy and stability of cavity length are affected by material defects, component fatigue, and environmental changes, resulting in center wavelength drift. Therefore, the center wavelength of the filter must be calibrated. Traditional calibration methods require a stable continuous spectrum light source and a spectrometer or interferometer with a resolution higher than that of the filter, which are large and expensive. Currently, the common calibration method uses the sun as a standard light source, but this is limited by variations in the solar spectrum. We propose a novel method for calibrating the center wavelength of the FP filter. This method is simple in structure and does not require additional spectrometers or interferometers, avoiding the problem of singlewavelength 2π entanglement and overcoming the limitations of solar spectral calibration. The calibration accuracy is better than 0. 01 Å, fully meeting the requirements of the NVST twodimensional imaging spectrometer. Methods Our method for calibrating the center wavelength of the FP filter is based on the periodicity of the intensity modulation curve of the FP filter. First, we measure the unit step of the controller using the adjacent peaks of the singlewavelength intensity modulation curve. The 2π entanglement problem is then solved using the dualwavelength intensity modulation curve, and the corresponding relationship between cavity length and the controller value is established according to the positions between adjacent peaks of the dualwavelength intensity modulation curve. Using this relationship, we adjust the cavity length of the FP filter to the position of the center wavelength and calibrate the wavelength accurately. Finally, we conduct a calibration test on the FP filter and verify the feasibility of this calibration method. Results and Discussions We carry out the calibration test of the ET100- FS- 100 FP filter produced by IC Optical Systems Ltd. The test results show that when observing the Hα solar spectral line, adjusting the controller to the -101st step, and with a corresponding cavity length is 1001. 81121 μm, we obtain a center wavelength of 6562. 8 Å with an accuracy of 0. 0093 Å. Analysis reveals that the errors mainly arise from two sources: random errors, including laser intensity stability, detector photon noise, electronic noise of FP filter, digitaltoanalog conversion quantization error, and nonlinear error, resulting in a wavelength drift of 0. 0079 Å; and the numerical quantization error of the controller, resulting in a wavelength drift of 0. 0014 Å. The average unit step of the controller measured by the calibration test is 429. 8 pm, compared to the theoretical value of 488. 4 pm provided by the manufacturer, with a difference of 58. 6 pm. The discrepancy may be due to differences between the laboratory calibration environment and the manufacturer’s test environment. The changes of environmental factors (such as temperature, humidity, and pressure) cause the filter material to expand and contract, electronic components to become unstable, and the refractive index of air in the cavity to change, resulting in differences between measured and theoretical values. This indicates that the measured unit step of the controller is only valid for the environment at the time of calibration. Once the environment changes, recalibration is necessary, or temperature control of the FP filter is required to ensure its stability. Conclusions We propose a novel method for calibrating the center wavelength of the FP filter based on the periodicity of the intensity modulation curve. By using the dualwavelength intensity modulation curve of the filter, we measure the unit step of the controller, establish the corresponding relationship between cavity length and the controller value, and accurately calibrate the wavelength of the filter. A calibration system is built in the laboratory for testing the ET100- FS- 100 FP filter produced by IC Optical Systems Ltd. The results show that the unit step of controller is 429. 8 pm. When observing the Hα solar spectral line, adjusting the controller to the −101st step, and with a corresponding cavity length of 1001. 81121 μm, the center wavelength of the FP filter is 6562. 8 Å, with a calibration accuracy of 0. 0093 Å, fully meeting the requirements of the NVST twodimensional imaging spectrometer. Key words Fabry‒Perot filter; calibration; cavity length; center wavelength; dualwavelength intensity modulation |
资助项目 | 国家自然科学基金[12073074]; 国家自然科学基金[11833010]; 国家重点研发计划[202103AD50013]; 云南省科技计划项目[202105AB160001]; 云南省科技人才与平台计划[202305AH340002]; 云南省兴滇英才人才培养计划[202305AS350029] |
项目资助者 | 国家自然科学基金[12073074, 11833010] ; 国家重点研发计划[202103AD50013] ; 云南省科技计划项目[202105AB160001] ; 云南省科技人才与平台计划[202305AH340002] ; 云南省兴滇英才人才培养计划[202305AS350029] |
语种 | 中文 |
学科领域 | 物理学 ; 光学 ; 天文学 ; 天文学其他学科 ; 机械工程 ; 仪器仪表技术 ; 光学技术与仪器 ; 天文仪器 |
ISSN | 0253-2239 |
引用统计 | |
文献类型 | 期刊论文 |
版本 | 出版稿 |
条目标识符 | http://ir.ynao.ac.cn/handle/114a53/27986 |
专题 | 天文技术实验室 |
作者单位 | 1.中国科学院云南天文台; 2.中国科学院大学 |
第一作者单位 | 中国科学院云南天文台 |
推荐引用方式 GB/T 7714 | 朱瑾,马琳,金振宇,等. 基于双波长强度调制的法布里-珀罗滤光器中心波长定标[J]. 光学学报/Acta Optica Sinica,2024,44(24):185-192. |
APA | 朱瑾,马琳,金振宇,&穆恒宇.(2024).基于双波长强度调制的法布里-珀罗滤光器中心波长定标.光学学报/Acta Optica Sinica,44(24),185-192. |
MLA | 朱瑾,et al."基于双波长强度调制的法布里-珀罗滤光器中心波长定标".光学学报/Acta Optica Sinica 44.24(2024):185-192. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
基于双波长强度调制的法布里-珀罗滤光器中(1219KB) | 期刊论文 | 出版稿 | 开放获取 | CC BY-NC-SA | 浏览 请求全文 |
个性服务 |
推荐该条目 |
保存到收藏夹 |
查看访问统计 |
导出为Endnote文件 |
谷歌学术 |
谷歌学术中相似的文章 |
[朱瑾]的文章 |
[马琳]的文章 |
[金振宇]的文章 |
百度学术 |
百度学术中相似的文章 |
[朱瑾]的文章 |
[马琳]的文章 |
[金振宇]的文章 |
必应学术 |
必应学术中相似的文章 |
[朱瑾]的文章 |
[马琳]的文章 |
[金振宇]的文章 |
相关权益政策 |
暂无数据 |
收藏/分享 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论